Evaluation of a commercial Monte Carlo dose calculation algorithm for electron treatment planning

评估用于电子治疗计划的商业蒙特卡罗剂量计算算法

阅读:1

Abstract

The RayStation treatment planning system implements a Monte Carlo (MC) algorithm for electron dose calculations. For a TrueBeam accelerator, beam modeling was performed for four electron energies (6, 9, 12, and 15 MeV), and the dose calculation accuracy was tested for a range of geometries. The suite of validation tests included those tests recommended by AAPM's Medical Physics Practice Guideline 5.a, but extended beyond these tests in order to validate the MC algorithm in more challenging geometries. For MPPG 5.a testing, calculation accuracy was evaluated for square cutouts of various sizes, two custom cutout shapes, oblique incidence, and heterogenous media (cork). In general, agreement between ion chamber measurements and RayStation dose calculations was excellent and well within suggested tolerance limits. However, this testing did reveal calculation errors for the output of small cutouts. Of the 312 output factors evaluated for square cutouts, 20 (6.4%) were outside of 3% and 5 (1.6%) were outside of 5%, with these larger errors generally being for the smallest cutout sizes within a given applicator. Adjustment of beam modeling parameters did not fix these calculation errors, nor does the planning software allow the user to input correction factors as a function of field size. Additional validation tests included several complex phantom geometries (triangular nose phantom, lung phantom, curved breast phantom, and cortical bone phantom), designed to test the ability of the algorithm to handle high density heterogeneities and irregular surface contours. In comparison to measurements with radiochromic film, RayStation showed good agreement, with an average of 89.3% pixels passing for gamma analysis (3%/3mm) across four phantom geometries. The MC algorithm was able to accurately handle the presence of irregular surface contours (curved cylindrical phantom and a triangular nose phantom), as well as heterogeneities (cork and cortical bone).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。