A novel method for monitoring the constancy of beam path accuracy in CyberKnife

一种用于监测CyberKnife光束路径精度稳定性的新方法

阅读:1

Abstract

The aim of current work was to present a novel evaluation procedure implemented for checking the constancy of beam path accuracy of a CyberKnife system based on ArcCHECK. A tailor-made Styrofoam with four implanted fiducial markers was adopted to enable the fiducial tracking during beam deliveries. A simple two-field plan and an isocentric plan were created for determining the density override of ArcCHECK in MultiPlan and the constancy of beam path accuracy respectively. Correlation curves for all diodes involved in the study were obtained by analyzing the dose distributions calculated by MultiPlan after introducing position shifts in anteroposterior, superoinferior, and left-right directions. The ability of detecting systematic position error was also evaluated by changing the position of alignment center intentionally. The one standard deviation (SD) result for reproducibility test showed the RMS of 0.054 mm and the maximum of 0.263 mm, which was comparable to the machine self-test result. The mean of absolute value of position errors in the constancy test was measured to 0.091 mm with a SD of 0.035 mm, while the root-mean-square was 0.127 mm with a SD of 0.034 mm. All introduced systematic position errors range from 0.3 to 2 mm were detected successfully. Efficient method for evaluating the constancy of beam path accuracy of CyberKnife has been developed and proven to be sensitive enough for detecting a systematic drift of robotic manipulator. Once the workflow is streamlined, our proposed method will be an effective and easy quality assurance procedure for medical physicists.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。