Beam focal spot position: The forgotten linac QA parameter. An EPID-based phantomless method for routine Stereotactic linac QA

光束焦点位置:被遗忘的直线加速器 QA 参数。用于常规立体定向直线加速器 QA 的基于 EPID 的无模型方法

阅读:1

Abstract

Modern day Stereotactic treatments require high geometric accuracy of the delivered treatment. To achieve the required accuracy the IGRT imaging isocenter needs to closely coincide with the treatment beam isocenter. An influence on this isocenter coincidence and on the spatial positioning of the beam itself is the alignment of the treatment beam focal spot with collimator rotation axis. The positioning of the focal spot is dependent on the linac beam steering and on the stability of the monitor chamber and beam steering servo system. As such, there is the potential for focal spot misalignment and this should be checked on a regular basis. Traditional methods for measuring focal spot position are either indirect, inaccurate, or time consuming and hence impractical for routine use. In this study a novel, phantomless method has been developed using the EPID (Electronic Portal Imaging Device) that utilizes the different heights of the MLC and jaws. The method has been performed on four linear accelerators and benchmarked against an alternate ion chamber-based method. The method has been found to be reproducible to within ±0.012 mm (1 SD) and in agreement with the ion chamber-based method to within 0.001 ± 0.015 mm (1 SD). The method could easily be incorporated into a departmental routine linac QA (Quality Assurance) program.