Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance

用于VMAT调试和质量保证的时间分辨光束对称性测量

阅读:1

Abstract

In volumetric-modulated arc therapy (VMAT) treatment delivery perfect beam symmetry is assumed by the planning system. This study aims to test this assumption and present a method of measuring time-resolved beam symmetry measurement during a VMAT delivery that includes extreme variations of dose rate and gantry speed. The Sun Nuclear IC Profiler in gantry mount was used to measure time-resolved in-plane and cross-plane profiles during plan delivery from which symmetry could be determined. Time-resolved symmetry measurements were performed throughout static field exposures at cardinal gantry angles, conformal arcs with constant dose rate and gantry speed, and during a VMAT test plan with gantry speed and dose rate modulation. Measurements were performed for both clockwise and counterclockwise gantry rotation and across four Varian 21iX lin-acs. The symmetry was found to be generally constant throughout the static field exposures to within 0.3% with an exception on one linac of up to 0.7%. Agreement in symmetry between cardinal angles was always within 1.0% and typically within 0.6%. During conformal arcs the results for clockwise and counterclockwise rotation were in agreement to within 0.3%. Both clockwise and counterclockwise tended to vary in similar manner by up to 0.5% during arc consistent with the cardinal gantry angle static field results. During the VMAT test plan the symmetry generally was in agreement with the conformal arc results. Greater variation in symmetry was observed in the low-dose-rate regions by up to 1.75%. All results were within clinically acceptable levels using the tolerances of NCS Report 24 (2015).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。