Effect of interfractional shoulder motion on low neck nodal targets for patients treated using volumetric-modulated arc therapy (VMAT)

肩部运动对采用容积调强弧形治疗(VMAT)的患者颈下淋巴结靶区的影响

阅读:1

Abstract

VMAT is an important tool in the treatment of head and neck cancers, many of which also require treatment to the supraclavicular lymph nodes. However, full VMAT arcs treating this nodal region necessarily cause entrance beam to pass through patients' shoulders. Thus, interfractional variations in shoulder position may cause unwanted dose perturbations. To assess this possibility, six patients undergoing treatment at our institution for head and neck cancers with associated supraclavicular lymph node treatment were imaged with in-room CT-on-rails during the course of their treatments. This allowed for the establishment of a true record of the actual shoulder position during selected treatment fractions. Then, a full VMAT plan and a plan with VMAT arcs superior to the shoulder and a static anteroposterior field inferiorly were copied onto the patients' weekly image sets. The average one-dimensional shoulder motion was generally within 10 mm of the simulated position, with some notable exceptions. The standard deviation in week-to-week shoulder position relative to simulation was 4.3 mm and 4.2 mm in the SI and AP dimensions, respectively. The average nodal target mean dose across all fractions sampled was within 5% of planned for all patients and both plans. Similarly, the average D95 for the nodal target was within 5% of planned across all fractions sampled, with the single exception of the full VMAT plan for one patient. In most cases, the standard deviation in both target mean dose and D95 was smaller with the VMAT+static AP field plan than it was with the full VMAT plan.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。