Noncoplanar verification: a feasibility study using Philips' Pinnacle3 treatment planning system

非共面验证:使用飞利浦 Pinnacle3 治疗计划系统的可行性研究

阅读:1

Abstract

Noncoplanar fields are normally used to improve the dose conformity of the target while sparing organs at risk. One of the methods to verify the dose distribution from the noncoplanar fields is by comparing their planar dose distributions from the treatment planning system (TPS) and the measured ones; for example, using film or electronic portal imaging devices (EPID). The planar dose distributions of the measurement tools, that are normally perpendicular to the central axis of the beam, can be calculated by creating special structures to mimic them in the TPS. With TPS commercially available today, however, it is not easy to create these special structures, especially in the noncoplanar configuration. For this work, we have written in-house scripts in the Pinnacle(3) TPS that can create the structures and define them as virtual planes. These virtual planes can be generated for any arbitrary gantry and couch angles, as well as source to planar distance, so that the planar dose maps at these planes can be computed. Two independent quality assurance (QA) tools were used to validate the planar dose distributions computed using the scripts for three open fields and one IMRT field at several different couch angles. The absolute planar dose patterns measured by the QA tools for all fields at all couch angles were found to be in good agreement, more than 95% (gamma criteria 3% delta dose and 3 mm distance to agreement), with the calculated ones by TPS. The results of this feasibility study can be valuable either for pretreatment dose verification or for in vivo dosimetry that directly implements the planar dose distributions from the TPS, particularly for the noncoplanar fields.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。