The sensitivity of ArcCHECK-based gamma analysis to manufactured errors in helical tomotherapy radiation delivery

基于 ArcCHECK 的伽马分析对螺旋断层放射治疗过程中制造误差的敏感性

阅读:1

Abstract

Three-dimensional measurement arrays are an efficient means of acquiring a distribution of data for patient plan delivery QA. However, the tie between plan integrity and traditional gamma-based analysis of these data are not clear. This study explores the sensitivity of such analysis by creating errors in Helical Tomotherapy delivery and measuring the passing rates with an ArcCHECK cylindrical diode array. Errors were introduced in each of the couch speed, leaf open time, and gantry starting position in increasing magnitude while the resulting gamma passing rates were tabulated. The error size required to degrade the gamma passing rate to 90% or below was on average a 3% change in couch speed, 5° in gantry synchronization, or a 5 ms in leaf closing speed for a 3%/3 mm Van Dyk gamma analysis. This varied with plan type, with prostate plans exhibiting less sensitivity than head and neck plans and with gamma analysis criteria, but in all cases the error magnitudes were large compared to actual machine tolerances. These findings suggest that the sensitivity of ArcCHECK-based gamma analysis to single-mode errors in tomotherapy plans is dependent upon plan and analysis type and at traditional passing thresholds unable to detect small defects in the plan.