Evaluation of new data processing algorithms for planar gated ventriculography (MUGA)

平面门控心室造影(MUGA)新数据处理算法的评估

阅读:1

Abstract

Before implementing one of two new LVEF radionuclide gated ventriculogram (MUGA) systems, the results from 315 consecutive parallel patient studies were evaluated. Each gamma-camera acquisition was simultaneously processed by semi-automatic Medasys Pinnacle and by fully-automatic and semi-automatic Philips nuclear medicine computer systems. The Philips systems yielded LVEF results within +/- 5 LVEF percentage points of the Medasys system in fewer than half of the studies. The remaining values were higher or lower than those from the long-used Medasys system. These differences might have changed cancer patient chemotherapy clinical decisions in 54 cases (17% of studies) for one of the systems. As a result, our institution elected not to implement either new system.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。