On the need for monitor unit calculations as part of a beam commissioning methodology for a radiation treatment planning system

关于将监测单位计算作为放射治疗计划系统束流调试方法的一部分的必要性

阅读:1

Abstract

This paper illustrates the need for validating the calculation of monitor units as part of the process of commissioning a photon beam model in a radiation treatment planning system. Examples are provided in which this validation identified subtle errors, either in the dose model or in the implementation of the dose algorithm. These errors would not have been detected if the commissioning process only compared relative dose distributions. A set of beam configurations, with varying field sizes, source-to-skin distances, wedges, and blocking, were established to validate monitor unit calculations for two different beam models in two different radiation treatment planning systems. Monitor units calculated using the treatment planning systems were compared with monitor units calculated from point dose calculations from tissue-maximum ratio (TMR) tables. When discrepancies occurred, the dose models and the code were analyzed to identify the causes of the discrepancies. Discrepancies in monitor unit calculations were both significant (up to 5%) and systematic. Analysis of the dose computation software found: (1) a coordinate system transformation error, (2) mishandling of dose-spread arrays, (3) differences between dose calculations in the commissioning software and the planning software, and (4) shortcomings in modeling of head scatter. Corrections were made in the beam calculation software or in the data sets to overcome these discrepancies. Consequently, we recommend incorporating validation of monitor unit calculations as part of a photon beam commissioning process.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。