Automation of a spectrophotometric method for measuring L -carnitine in human blood serum

自动化分光光度法测定人血清中左旋肉碱

阅读:1

Abstract

A spectrometric method for the determination of L-carnitine has been developed based on the reaction of the 5,5' dithiobis-(2-nitrobenzoic) acid (DTNB) and adapted to a Technicon RA-2000 automatic analyser Química Farmacéutica Bayer, S.A.). The detection limit of the method is 13.2 mumol/l, with a measurement interval ranging from 30 to 320 mumoll1. Imprecision and accuracy are good even at levels close to the detection limit (coeffcient of variation of 5.4% for within-run imprecision for a concentration of 35 mumol/l). A good correlation was observed between the method studied and the radiometric method. The method evaluated has suffcient analytical sensitivity to diagnose carnitine deficiencies. The short time period required for sample processing (30 samples in 40min), the simple methodology and apparatus, the ease of personnel training and the low cost of the reagents make this method a good alternative to the classical radiometric method for evaluating serum L-carnitine in clinical laboratories without radioactive installations.