Characteristics Performance of Newly Developed Parallel Plate Chamber in Electron Beams

新型平行板室在电子束中的特性及性能

阅读:1

Abstract

BACKGROUND: To investigate the dosimetric performance of newly developed parallel plate chamber in electron beams. MATERIALS AND METHODS: Rosalina Instruments India Private Limited (Mumbai, Maharashtra, India) has designed and fabricated PRATT2 parallel plate chamber. The various dosimetric characteristics, including pre- and post-irradiation leakage, stability, polarity effect, chamber response with bias voltage, dose linearity, dose rate effect, and chamber absorbed dose calibration, were performed for the developed chamber. The electron beam energies of 4, 6, 8, and 15MeV were used in this study. RESULTS: The pre- and post-irradiation leakage of the developed chamber was within the acceptable limit. The chamber shows good stability over the electron beams used in this study. The maximum error in polarity effect was 0.7% for the developed chamber. The chamber shows the good linear response with dose, and its response is independent of the dose rate for all electron beams. The beam quality correction factor (k(Q, Q0)) was determined for the all electron beam energies, which was used for determination absorbed dose in electron beams. DISCUSSION: The developed parallel plate chamber (PRATT2) is suitable for dosimetry of electron beams in radiotherapy. The chamber is cost effective and shows precise and reproducible response. The study carried out confirms that the newly designed and fabricated ion chamber can be used in the measurement of absorbed dose for therapeutic electron beams.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。