Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study

利用三种不同的锥形束计算机断层扫描设备进行颌面部诊断的散射辐射分布:一项研究

阅读:1

Abstract

This study aimed to estimate scattered radiation and its spatial distribution around three cone-beam computed tomography (CBCT) devices, in order to determine potential positions for an operator to stand if they needed to be inside the CBCT room. The following devices were tested: Morita Accuitomo (CBCT1), Newtom Giano HR (CBCT2), Newtom VGi (CBCT3). Scattered radiation measurements were performed using different kVp, mA, and Field of View (FOV) options. An anthropomorphic phantom (NATHANIA) was placed inside the X-ray gantry to simulate clinical conditions. Scattered measurements were taken with the Inovision model 451P Victoreen ionization chamber once placed at fixed distances from each irradiation isocenter, away from the primary beam. A statistically significant (p < 0.001) difference was found in the mean value of the scattered radiation estimations between the CBCT devices. Scattered radiation was reduced with a different rate for each CBCT device as distance was increased. For CBCT1 the reduction was 0.047 μGy, for CBCT2 it was 0.036 μGy, and for CBCT3 it was 0.079 μGy, for every one meter from the X-ray gantry. Therefore, at certain distances from the central X-ray, the scattered radiation was below the critical level of 1 mGy, which is defined by the radiation protection guidelines as the exposure radiation limit of the general population. Consequently, an operator could stay inside the room accompanying the patient being scanned, if necessary.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。