Artificial neural network analysis of the oxygen saturation signal enables accurate diagnostics of sleep apnea

利用人工神经网络分析氧饱和度信号可以准确诊断睡眠呼吸暂停

阅读:1

Abstract

The severity of obstructive sleep apnea (OSA) is classified using apnea-hypopnea index (AHI). Accurate determination of AHI currently requires manual analysis and complicated registration setup making it expensive and labor intensive. Partially for these reasons, OSA is a heavily underdiagnosed disease as only 7% of women and 18% of men suffering from OSA have diagnosis. To resolve these issues, we introduce an artificial neural network (ANN) that estimates AHI and oxygen desaturation index (ODI) using only the blood oxygen saturation signal (SpO2), recorded during ambulatory polygraphy, as an input. Therefore, hypopneas associated only with an arousal were not considered in this study. SpO2 signals from 1692 patients were used for training and 99 for validation. Two test sets were used consisting of 198 and 1959 patients. In the primary test set, the median absolute errors of ANN estimated AHI and ODI were 0.78 events/hour and 0.68 events/hour respectively. Based on the ANN estimated AHI and ODI, 90.9% and 94.4% of the test patients were classified into the correct OSA severity category. In conclusion, AHI and ODI can be reliably determined using neural network analysis of SpO2 signal. The developed method may enable a more affordable screening of OSA.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。