Performance Characteristics of a New Generation 148-cm Axial Field-of-View uMI Panorama GS PET/CT System with Extended NEMA NU 2-2018 and EARL Standards

新一代148厘米轴向视野uMI全景GS PET/CT系统的性能特征(符合扩展的NEMA NU 2-2018和EARL标准)

阅读:1

Abstract

The uMI Panorama GS PET/CT system is a new long-axial-field-of-view scanner featuring high sensitivity, time-of-flight (TOF) resolution, spatial resolution, and count rate performance. The aim of this study is to assess the PET system on the basis of the National Electrical Manufacturers Association (NEMA) NU 2-2018 and European Association of Nuclear Medicine Research Limited (EARL) standards. Methods: Spatial resolution, count rate performance, sensitivity, accuracy, image quality, TOF resolution, and coregistration accuracy were evaluated following the NEMA NU 2-2018 standard. Additional experiments included energy resolution, 200-cm-long line sources for sensitivity, a 175-cm-long scatter phantom for count rate and TOF resolution, as well as the compliance with the EARL guideline. Moreover, an (18)F-FDG PET patient study was reconstructed with various frame durations. Results: The PET system achieved sub-3-mm transaxial and axial spatial resolutions at a 1-cm radial offset. The sensitivities with the 70-cm-long and 200-cm-long line sources were observed to be 176.3 and 90.8 kcps/MBq, respectively, at the center of the field of view. The noise-equivalent count rates (NECRs) of the 70-cm-long and 175-cm-long scatter phantoms were measured to be 3.35 Mcps at 57.57 kBq/mL and 2.24 Mcps at 33.27 kBq/mL, respectively. The TOF resolutions for both phantoms were approximately 189 ps at 5.3 kBq/mL and lower than 200 ps below the NECR peaks. The absolute count rate errors of all 34 acquisitions were less than 3% below the NECR peak for the 70-cm-long scatter phantom. With the standard NEMA image quality phantom experiment, the contrast recovery coefficient varied from 68.17% to 94.20% and the background variabilities were all below 2%. The maximum PET/CT coregistration error was 1.33 mm. Regarding EARL compliance, the gaussian filter of 5-mm full width at half maximum could produce acceptable images. The patient data demonstrate visually satisfactory image quality with short frames (less than 1 min). Conclusion: The uMI Panorama GS exhibits spatial resolution and TOF resolution similar to those of the uMI Panorama system (35-cm axial field of view), despite the extended axial field of view. The 148-cm axial coverage, sub-200-ps TOF resolution, high sensitivity, and count rate performances are expected to yield superior image quality and offer new opportunities for various clinical applications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。