Electron microscopy procedure influences detection of rotaviruses

电子显微镜操作会影响轮状病毒的检测

阅读:1

Abstract

Technical parameters of electron microscope staining procedures (type of stain, pH of stain, and time of staining) influence particle integrity for three groups of rotaviruses. Simian rotavirus SA11 (group A), Chinese adult diarrhea rotavirus and porcine rotavirus-like agent (group B), and porcine pararotavirus (group C) were tested. All rotavirus strains were quite stable in uranyl acetate and phosphotungstic acid at pH 4.5 and relatively stable in ammonium molybdate. However, staining with phosphotungstic acid at higher pH values with increased staining time yielded a reduction in the number of particles and particles that were broken or degraded to single-shelled particles or core particles. The different staining procedures were also tested in immunoelectron microscopy experiments. Antibody molecules bound to rotavirus particles were observed clearly only with phosphotungstic acid staining and not with uranyl acetate. We therefore recommend that uranyl acetate and phosphotungstic acid at pH 4.5 be used for negative staining of rotaviruses; phosphotungstic acid at pH 4.5 is optimal for immunoelectron microscopy. These technical points may be critical for rotavirus detection and are important for studies pertaining to the epidemiology and clinical importance of the non-group A rotaviruses.