Digital breast tomosynthesis: studies of the effects of acquisition geometry on contrast-to-noise ratio and observer preference of low-contrast objects in breast phantom images

数字乳腺断层合成:采集几何形状对乳腺模型图像中低对比度物体对比噪声比和观察者偏好的影响研究

阅读:1

Abstract

The effect of acquisition geometry in digital breast tomosynthesis was evaluated with studies of contrast-to-noise ratios (CNRs) and observer preference. Contrast-detail (CD) test objects in 5 cm thick phantoms with breast-like backgrounds were imaged. Twelve different angular acquisitions (average glandular dose for each ~1.1 mGy) were performed ranging from narrow angle 16° with 17 projection views (16d17p) to wide angle 64d17p. Focal slices of SART-reconstructed images of the CD arrays were selected for CNR computations and the reader preference study. For the latter, pairs of images obtained with different acquisition geometries were randomized and scored by 7 trained readers. The total scores for all images and readings for each acquisition geometry were compared as were the CNRs. In general, readers preferred images acquired with wide angle as opposed to narrow angle geometries. The mean percent preferred was highly correlated with tomosynthesis angle (R = 0.91). The highest scoring geometries were 60d21p (95%), 64d17p (80%), and 48d17p (72%); the lowest scoring were 16d17p (4%), 24d9p (17%) and 24d13p (33%). The measured CNRs for the various acquisitions showed much overlap but were overall highest for wide-angle acquisitions. Finally, the mean reader scores were well correlated with the mean CNRs (R = 0.83).