Atom Probe Mass Spectrometry of Uranium Isotopic Reference Materials

铀同位素参考物质的原子探针质谱分析

阅读:1

Abstract

Atom probe tomography (APT)-based isotopic analyses are becoming increasingly attractive for analysis applications requiring small volumes of material and sub-micrometer length scales, such as isotope geochemistry, nuclear safety, and materials science. However, there is an open question within the atom probe community as to the reliability of atom probe isotopic and elemental analyses. Using our proposed analysis guidelines, in conjunction with an empirical calibration curve and a machine learning-based adaptive peak fitting algorithm, we demonstrate accurate and repeatable uranium isotopic analyses, via atom probe mass spectrometry, on U(3)O(8) isotopic reference materials. By using isotopic reference materials, each measured isotopic abundance value could be directly compared to a known certified reference value to permit a quantitative statement of accuracy. The isotopic abundance measurements for (235)U and (238)U in each individual APT sample were consistently within ±1.5% relative to the known reference values. The accuracy and repeatability are approaching values consistent with measurements limited primarily by Poisson counting statistics, i.e., the number of uranium atoms recorded.