Real-Time Heart Rate Detection Method Based on 77 GHz FMCW Radar

基于77 GHz FMCW雷达的实时心率检测方法

阅读:1

Abstract

This paper proposes a real-time heart rate detection method based on 77 GHz FMCW radar. Firstly, the method establishes a new motion model according to respiratory and heartbeat rules, and extracts the motion signals of the chest and the abdomen; then, the random body motion (RBM) signal is eliminated by a combination of polynomial fitting and recursive least squares (RLS) adaptive filtering; lastly, multi-detection-point adaptive harmonics cancellation (AHC) is used to eliminate respiratory harmonics. In addition, the method introduces a spectrum analysis algorithm based on linear predictive coding (LPC). The experimental results show that the method can effectively eliminate the RBM signal and respiratory harmonics, and that the average real-time heart rate detection error rate is 2.925%.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。