Nitrate Sensor with a Wide Detection Range and High Stability Based on a Cu-Modified Boron-Doped Diamond Electrode

基于铜修饰硼掺杂金刚石电极的宽检测范围、高稳定性硝酸盐传感器

阅读:1

Abstract

This paper describes an electrochemical sensor based on a Cu-modified boron-doped diamond (BDD) electrode for the detection of nitrate-contaminated water. The sensor utilizes the catalytic effect of copper on nitrate and the stability of the BDD electrode. By optimizing the electrolyte system, the linear detection range was expanded, allowing the sensor to detect highly concentrated nitrate samples up to 100 mg/L with a low detection limit of 0.065 mg/L. Additionally, the stability of the sensor was improved. The relative standard deviation of the current responses during 25 consecutive tests was only 1.03%. The wide detection range and high stability of the sensor makes it suitable for field applications and the on-site monitoring of nitrate-contaminated waters.