Abstract

Objectives: This study aimed to compare the ex vivo performance of two rotary nickel-titanium (NiTi) systems with similar designs but manufactured from martensitic and austenitic alloys, the One Curve (OC) and One Shape (OS) rotary endodontic files, respectively. Methods: Forty separate mesial canals of 20 extracted mandibular molars were scanned using micro-computed tomography (CT), which were divided into 2 groups and instrumented with OC and OS, respectively. Post-instrumentation micro-CT scans were evaluated using validated computer algorithms to compare changes in canal thickness, surface area, structure model index (SMI) scores, volume of removed dentin, percentage of untreated canal surface, percentage of curvature straightening, and the amount of canal transportation. Results: Both files led to significant changes in the basic root canal geometry, with no preparation errors and no statistically significant differences. However, OC treatment resulted in significantly less curvature straightening (17.30%; 10.77%) (independent samples t test, p < 0.05) and less apical transportation (55.11 µm; 33.15 µm) (Mann-Whitney U, test p < 0.05) compared to OS treatment. Transportation values in the middle and coronal thirds were statistically similar (independent sample t-test, p > 0.05). OC treatment produced significantly less straightening and less apical transportation than OS.