Influence of injection rates of calibrating standard solution on monitoring pulse indicator continuous cardiac output

校准标准溶液注射速率对脉搏指示器连续心输出量监测的影响

阅读:1

Abstract

OBJECTIVE: This study aimed to investigate the influence of injection rates of calibrating standard solution on monitoring pulse indicator continuous cardiac output (PICCO, made in Germany), and thereby to provide significant references for clinical practice. METHODS: A total of 108 critical patients in stroke intensive care unit were identified. All these participants received transesophageal cardiac color Doppler ultrasound, and within 15 min PICCO equipment was utilized to monitor the relevant parameters, by means of 0 °C calibrating standard solution, and the injection speeds were 2-4, 5-7, and 8-10 s. Besides, the monitoring indicators were as follows, cardiac index, global ejection fraction, global end diastolic volume index. The potential correlations were evaluated between PICCO and transesophageal cardiac color Doppler ultrasound. RESULTS: All the data was available, and the monitored parameters of PICOO at 2-4, 5-7, and 8-10 s were positively correlated with the parameters obtained from transesophageal cardiac color Doppler ultrasound (P < 0.05). Specially, it is worth emphasizing that the best correlation between them could be provided when the injection rate was 2-4 s. CONCLUSION: When the injection rate at 2-4 s, the parameters obtained by PICOO were much closer to that of transesophageal cardiac color Doppler ultrasound. Furthermore, the parameters of PICOO obtained at 2-4 s could better reflect cardiac function of patients.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。