Estimation of Spatiotemporal Gait Parameters in Walking on a Photoelectric System: Validation on Healthy Children by Standard Gait Analysis

利用光电系统估计行走时空步态参数:通过标准步态分析对健康儿童进行验证

阅读:1

Abstract

The use of stereophotogrammetry systems is challenging when targeting children's gait analysis due to the time required and the need to keep physical markers in place. For this reason, marker-less photoelectric systems appear to be a solution for accurate and fast gait analysis in youth. The aim of this study is to validate a photoelectric system and its configurations (LED filter setting) on healthy children, comparing the kinematic gait parameters with those obtained from a three-dimensional stereophotogrammetry system. Twenty-seven healthy children were enrolled. Three LED filter settings for the OptoGait were compared to the BTS P6000. The analysis included the non-parametric 80% limits of agreement and the intraclass correlation coefficient (ICC). Additionally, normalised limits of agreement and bias (NLoAs and Nbias) were compared to the clinical experience of physical therapists (i.e., assuming an error lower than 5% is acceptable). ICCs showed excellent consistency for most of the parameters and filter settings; NLoAs varied between 1.39% and 12.62%. An inverse association between the number of LEDs for filter setting and the bias values was also observed. Observations confirm the validity of the OptoGait system for the evaluation of spatiotemporal gait parameters in children.