Reliability and accuracy of intraoperative fluoroscopy assessment of acetabular cup anteversion in supine direct anterior approach total hip arthroplasty

仰卧位直接前入路全髋关节置换术中术中透视评估髋臼杯前倾角的可靠性和准确性

阅读:1

Abstract

Poor implantation positioning of hip prostheses is considered the primary factor affecting postoperative joint wear. Cup anteversion in direct anterior approach (DAA) total hip arthroplasty (THA) is often excessive. Intraoperative fluoroscopy (IF) are effective for improving implant placement accuracy. This study aimed to analyze IF's reliability and accuracy in assessing intraoperative anteversion. Sixty-two consecutive hips underwent primary THA utilizing DAA alongside IF for cup placement. Intraoperative anteversion was measured using IF images, while postoperative CT and standard anteroposterior (AP) radiographs were used to calculate true anteversion component angles. Differences and correlations between intraoperative and true anteversions were analyzed, and intraclass correlation coefficients (ICC) determined the inter- and intra-observer reliabilities. Excellent intra- and inter-observer reliabilities were observed for all radiographic and CT methods (ICC > 0.9). Strong correlations (PCC > 0.6) existed between anteversion measured on IF image and postoperative CT and AP pelvic measurements. Intraoperative anteversion measured on IF images (16.8 ± 3.2°) was smaller than anteversion measured postoperatively on AP X-rays (21.3 ± 4.7°, P < 0.001) and CT (22.0 ± 4.9°, P < 0.001), with average differences of 4.5°and 5.3°, respectively. Under several influencing factors, the accuracy of IF in assessing cup anteversion in DAA-THA may be limited. However, this still requires large-sample experiments for verification.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。