Evaluation of cost-effective system designs for long axial field-of-view PET scanners

长轴向视野PET扫描仪经济高效系统设计的评估

阅读:1

Abstract

Objective. Current commercial positron emission tomography (PET) scanners have excellent performance and diagnostic image quality primarily due to improvements in scanner sensitivity and time-of-flight (TOF) resolution. Recent years have seen the development of total-body PET scanners with longer axial field-of-view (AFOV) that increase sensitivity for single organ imaging, and also image more of the patient in a single bed position thereby enabling multi-organ dynamic imaging. While studies have shown significant capabilities of these systems, cost will be a major factor in their widespread adoption in the clinic. Here we evaluate alternative designs that achieve many advantages of long AFOV PET while utilizing cost-effective detector hardware.Approach. We utilize Monte Carlo simulations and clinically relevant lesion detectability metric to study the impact of scintillator type lutetium oxyorthosilicate or bismuth germanate (LSO or BGO), scintillator thickness (10-20 mm), and TOF resolution on resultant image quality in a 72 cm long scanner. Detector TOF resolution was varied based on current scanner performance, as well as expected future performance from detector designs that currently hold most promise for scaling into a scanner.Main results. Results indicate that BGO is competitive with LSO (both 20 mm thick) if we assume that it uses TOF (e.g. Cerenkov timing with 450 ps fwhm and Lorentzian distribution) and the LSO scanner has TOF resolution similar to the latest PMT-based scanners (∼500-650 ps). Alternatively, a system using 10 mm thick LSO with 150 ps TOF resolution can also provide similar performance. Both these alternative systems can provide cost savings (25%-33%) relative to a scanner using 20 mm LSO with ∼50% of effective sensitivity, but still 500%-700% higher than a conventional AFOV scanner.Significance. Our results have relevance to the development of long AFOV PET, where reduced cost of these alternative designs can provide wider accessibility for use in situations requiring imaging of multiple organs simultaneously.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。