Case of a Stuck Coil in the Final Stage of Stent-Assisted Coil Embolization

支架辅助弹簧圈栓塞术最后阶段弹簧圈卡住的病例

阅读:2

Abstract

OBJECTIVE: We report a case in which two coils became stuck in a microcatheter at the end of coil embolization for a cerebral aneurysm. CASE PRESENTATION: Two coils became stuck in the microcatheter at the final stage of stent-assisted coil embolization for an unruptured anterior communicating artery aneurysm. The rear end of a detached coil was near the tip of the microcatheter. The coil inserted next was pushed out of the microcatheter and pulled back into the microcatheter. Then, the rear end of the detached coil and the retracted coil meshed into the microcatheter, and became immobile. The microcatheter and these two coils were removed simultaneously, and coil embolization was finished. CONCLUSION: At the end of coil embolization, the filling rate is relatively high. Insertion of another coil and traction may cause the coils to become stuck in the microcatheter.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。