Focus variation due to near infrared laser in a confocal microscope

共聚焦显微镜中近红外激光引起的焦点变化

阅读:9
作者:Melissa Matrecitos-Avila, Remy Avila, Reinher Pimentel-Domínguez, Salvador Cuevas, Elisa Tamariz, Pablo Loza-Alvarez

Abstract

Focus precision and stability is crucial in confocal microscopy not only for image sharpness but also to avoid radiometric fluctuations that can wrongly be interpreted as variations of the fluorescence intensity in the sample. Here we report a focus variation provoked by a continuous wave laser of 810-nm wavelength introduced along the optical path of an inverted confocal microscope with an oil immersion ×60 objective. When the laser is turned on or off, the focus position drifts toward lower or high values of the vertical coordinate z, respectively. The maximum drift observed was 2.25 𝜇m for a laser power of 40 mW at the sample and over a 600-s exposure time. The temporal evolution of the focus position is well fitted by exponential curves that mimic temperature variations due to a heat source. Our analysis strongly suggests that the focus drift is due to heating of the immersion oil.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。