Diagnosis of and Early Revision Surgery for Biological Fixation Failure Due to Proximal-Distal Mismatch of Proximally Coated Tapered Cementless Stem

近端涂层锥形无骨水泥柄近远端不匹配导致生物固定失败的诊断和早期翻修手术

阅读:1

Abstract

The diagnosis of and decision-making for early revision surgery to treat failure of biological fixation with a proximally coated cementless stem are challenging. A 73-year-old woman was referred to our hospital with thigh pain 2 years after the initial total hip arthroplasty. Although a plain radiograph showed no signs indicating biological fixation failure, digital tomosynthesis showed a highly radiodense line along the proximal part, and bone scintigraphy showed uptake at the distal part. With the diagnosis of biological fixation failure due to the proximal-distal mismatch, the cementless stem was revised to a cemented stem, and the thigh pain was improved after the revision surgery. Digital tomosynthesis and bone scintigraphy can be helpful for the diagnosis of biological fixation failure due to proximal-distal mismatch.