Atrial tachycardia misinterpreted as ventricular high-rate event in a patient with the sick sinus disease and a DDD pacemaker

一名患有病态窦房结疾病并植入DDD起搏器的患者,其房性心动过速被误诊为室性心动过速。

阅读:2

Abstract

The device misinterpreted atrial high-rate episode (AHRE) as ventricular high-rate episode (VHRE), since consecutive atrial beats fell in the postventricular atrial blanking (PVAB) period. This resulted in the failure of auto mode switch (AMS). This was corrected by decreasing the PVAB and the patient was asymptomatic on follow-up.