High-speed and high-sensitivity parallel spectral-domain optical coherence tomography using a supercontinuum light source

采用超连续光源的高速高灵敏度并行谱域光学相干断层扫描

阅读:11
作者:Jessica Barrick, Ana Doblas, Michael R Gardner, Patrick R Sears, Lawrence E Ostrowski, Amy L Oldenburg

Abstract

The three most important metrics in optical coherence tomography (OCT) are resolution, speed, and sensitivity. Because there is a complex interplay between these metrics, no previous work has obtained the best performance in all three metrics simultaneously. We demonstrate that a high-power supercontinuum source, in combination with parallel spectral-domain OCT, achieves an unparalleled combination of resolution, speed, and sensitivity. This system captures cross-sectional images spanning 4 mm×0.5 mm at 1,024,000 lines/s with 2×14 μm resolution (axial×transverse) at a sensitivity of 113 dB. Imaging using the proposed system is demonstrated on highly differentiated human bronchial epithelial cells to capture and spatially localize ciliary dynamics.