Quantitative Detection of Cathepsin B Activity in Neutral pH Buffers Using Gold Microelectrode Arrays: Toward Direct Multiplex Analyses of Extracellular Proteases in Human Serum

使用金微电极阵列定量检测中性 pH 缓冲液中的组织蛋白酶 B 活性:对人血清中的细胞外蛋白酶进行直接多重分析

阅读:7
作者:Yang Song, Jestin Gage Wright, Morgan J Anderson, Sabari Rajendran, Zhaoyang Ren, Duy H Hua, Jessica E Koehne, M Meyyappan, Jun Li

Abstract

Proteases are critical signaling molecules and prognostic biomarkers for many diseases including cancer. There is a strong demand for multiplex bioanalytical techniques that can rapidly detect the activity of extracellular proteases with high sensitivity and specificity. This study demonstrates an activity-based electrochemical biosensor of a 3 × 3 gold microelectrode array for the detection of cathepsin B activity in human serum diluted in a neutral buffer. Proteolysis of ferrocene-labeled peptide substrates functionalized on 200 × 200 μm microelectrodes is measured simultaneously over the nine channels by AC voltammetry. The protease activity is represented by the inverse of the exponential decay time constant (1/τ), which equals to (kcat/KM)[CB] based on the Michaelis-Menten model. An enhanced activity of the recombinant human cathepsin B (rhCB) is observed in a low-ionic-strength phosphate buffer at pH = 7.4, giving a very low limit of detection of 8.49 × 10-4 s-1 for activity and 57.1 pM for the active rhCB concentration that is comparable to affinity-based enzyme-linked immunosorbent assay (ELISA). The cathepsin B presented in the human serum sample is validated by ELISA, which mainly detects the inactive proenzyme, while the electrochemical biosensor specifically measures the active cathepsin B and shows significantly higher decay rates when rhCB and human serum are activated. Analyses of the kinetic electrochemical measurements with spiked active cathepsin B in human serum provide further assessment of the protease activity in the complex sample. This study lays the foundation to develop the gold microelectrode array into a multiplex biosensor for rapid detection of the activity of extracellular proteases toward cancer diagnosis and treatment assessment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。