Dose reduction for kilovotage cone-beam computed tomography in radiation therapy

放射治疗中千伏锥形束计算机断层扫描的剂量降低

阅读:2

Abstract

Kilovotage cone-beam computed tomography (kV-CBCT) has shown potentials to improve the accuracy of a patient setup in radiotherapy. However, daily and repeated use of CBCT will deliver high extra radiation doses to patients. One way to reduce the patient dose is to lower mAs when acquiring projection data. This, however, degrades the quality of low mAs CBCT images dramatically due to excessive noises. In this work, we aim to improve the CBCT image quality from low mAs scans. Based on the measured noise properties of the sinogram, a penalized weighted least-squares (PWLS) objective function was constructed, and the ideal sinogram was then estimated by minimizing the PWLS objection function. To preserve edge information in the projection data, an anisotropic penalty term was designed using the intensity difference between neighboring pixels. The effectiveness of the presented algorithm was demonstrated by two experimental phantom studies. Noise in the reconstructed CBCT image acquired with a low mAs protocol was greatly suppressed after the proposed sinogram domain image processing, without noticeable sacrifice of the spatial resolution.