Structural differences and architectural features of two different polypropylene slings (TVT-O and I-STOP) have no impact on biocompatibility and tissue reactions

两种不同聚丙烯吊带(TVT-O 和 I-STOP)的结构差异和构造特征对生物相容性和组织反应没有影响。

阅读:1

Abstract

INTRODUCTION: To evaluate the impact of design features of the synthetic mid-urethral slings on tissue integrity and inflammatory responses. MATERIAL AND METHODS: In total 30 female Sprague-Dawley rats were implanted with type I monofilamentous, macroporous polypropylene meshes: Gynecare TVT-Obturator tape(®) (Ethicon Inc., Johnson & Johnson, Somerville, NJ, USA) and I-STOP(®) (CL Medical Inc., Lyon, France). All animal groups were sacrificed at set time intervals - 6 weeks, 3 months, 6 months, 9 months and 12 months - and the abdominal wall was harvested with mesh strips for histological evaluation. RESULTS: All mesh strips appeared to be well incorporated into the abdominal wall, and no signs of shrinkage was noticed. All specimens showed a thin/delicate, loose, fibrous interface between the synthetic graft plate and abdominal wall, along with mild inflammatory reactions from 6 weeks to 12 months. CONCLUSIONS: Both mesh brands induced comparable, minimal foreign body reactions and integrated well into the host tissues despite differences in architectural features. TVT-O(®) and I-STOP(®) evoked similar low-grade inflammatory responses up to 12 months in this animal model. Structural differences and architectural features of polypropylene slings used in this study have had no impact on tissue integrity and inflammatory responses.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。