Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug+Placido device

使用新型 Scheimpflug+Placido 设备进行前房生物测量和角膜厚度体积参数的单次操作、用户内重复性评估

阅读:2

Abstract

PURPOSE: To analyze single session, intrauser reliability of a Scheimpflug device for anterior chamber (AC) and corneal parameters. METHODS: In this observational study, 100 normal candidates underwent Scheimpflug analysis with Sirius 3D Rotating Scheimpflug Camera and Topography System (Costruzione Strumenti Oftalmici, Italy). Two scans in dark room conditions were performed by the same experienced user. The candidates were asked to keep both eyes closed for 5min before the scans. Exclusion criteria were previous ocular surgery, corneal scarring and anterior segment/posterior segment anomalies. Only the right eyes were used for the analysis. Both corneal (central, minimum, and apical thickness, volume, horizontal visible iris diameter, and apical curvature) and anterior chamber (volume, depth, angle, horizontal diameter) measurements were evaluated. RESULTS: There was no difference in the means of repeated measurements (p>0.05, ANOVA). Intraclass correlations between the measures were high and ranged from 0.995-0.997 for corneal to 0.964-0.997 for anterior chamber (AC) parameters. The precision of repeatability measures (1.96×Sw) was approximately 5μ for the central and minimum corneal thickness, 8μ for the apical corneal thickness, 0.06mm for AC (anterior chamber) depth and less than 2° for the AC angle. CONCLUSIONS: Sirius Scheimpflug system has high repeatability for both corneal and AC parameters in normal eyes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。