Accuracy of Leg Length and Offset Restoration in Femoral Pinless Navigation Compared to Navigation Using a Fixed Pin during Total Hip Arthroplasty

全髋关节置换术中,股骨无针导航与固定针导航在腿长和偏移恢复精度方面的比较

阅读:1

Abstract

Equalization of biomechanical differences is a major goal in total hip arthroplasty (THA). In the current study we compared the accuracy of restoring leg length and offset using imageless navigation with an osseous fixed pin to a femoral pinless device in 97 minimally invasive THAs through an anterolateral approach in the lateral decubitus position. Leg length and offset differences were evaluated on magnification-corrected radiographs by a blinded observer. A postoperative mean difference of -0.9 mm (95% CI -2.8 mm to 1.1 mm, p = 0.38) between pinless navigation and navigation with a fixed pin was observed for leg length and that of -2.4 mm (95% CI -3.9 mm to -0.9 mm, p = 0.002) was observed for offset, respectively. The number of patients with a residual difference below 5 mm after THA was higher if using a fixed pin than in pinless navigation for both leg length (98.2%, 54/55 to 50.0%, 21/42, p < 0.001) and offset (100.0%, 55/55 to 71.4%, 30/42, p < 0.001). Imageless navigation is a feasible method in intraoperative control of leg length and offset in minimally invasive THA. The use of pins fixed to the bone has a higher precision than pinless devices. This trial is registered with DRKS00000739.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。