Automated 3D-Printed Microfluidic Array for Rapid Nanomaterial-Enhanced Detection of Multiple Proteins

用于快速纳米材料增强检测多种蛋白质的自动化3D打印微流控阵列

阅读:1

Abstract

We report here the fabrication and validation of a novel 3D-printed, automated immunoarray to detect multiple proteins with ultralow detection limits. This low cost, miniature immunoarray employs electrochemiluminescent (ECL) detection measured with a CCD camera and employs touch-screen control of a micropump to facilitate automated use. The miniaturized array features prefilled reservoirs to deliver sample and reagents to a paper-thin pyrolytic graphite microwell detection chip to complete sandwich immunoassays. The detection chip achieves high sensitivity by using single-wall carbon nanotube-antibody conjugates in the microwells and employing massively labeled antibody-decorated RuBPY-silica nanoparticles to generate ECL. The total cost of an array is $0.65, and an eight-protein assay can be done in duplicate for $0.14 per protein with limits of detection (LOD) as low as 78-110 fg mL(-1) in diluted serum. The electronic control system costs $210 in components. Utility of the automated immunoarray was demonstrated by detecting an eight-protein prostate cancer biomarker panel in human serum samples in 25 min. The system is well suited to future clinical and point-of-care diagnostic testing and could be used in resource-limited environments.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。