Measuring CO2 and CH4 with a portable gas analyzer: Closed-loop operation, optimization and assessment

利用便携式气体分析仪测量二氧化碳和甲烷:闭环运行、优化和评估

阅读:1

Abstract

The use of cavity ring-down spectrometer (CRDS) based portable greenhouse gas analyzers (PGAs) in closed-loop configuration to measure small sample volumes (< 1 l) for CH4 and CO2 concentrations is increasing and offers certain advantages over conventional measurement methods in terms of speed as well as the ability to measure directly in field locations. This first systematic assessment of the uncertainties, problems and issues associated with achieving reliable and repeatable measurement with this technique presents the adaptation, measurement range, calibration and maintenance, accuracy and issues of efficient operation, for one example instrument. Regular open-loop calibration, a precise loop volume estimate, leak free system, and a high standard of injection practices are necessary for accurate results. For 100 μl injections, measured values ranging from 4.5 to 9 x104 ppm (CH4), and 1000 ppm to 1 x106 ppm (CO2) are possible with uncertainties ±5.9% and ±3.0%, respectively, beyond 100 ppm CH4 correction may be necessary. Uncertainty arising from variations water vapour content and atmospheric pressure are small (0.24% and -0.9% to +0.5%, respectively). With good practice, individual operator repeatability of 1.9% (CH4) and 2.48% (CO2) can be achieved. Between operator injection error was around 3% for both gases for four operators. Slow syringe plunger operation (> 1s) is recommended; generally delivered more (ca. 3-4%) sample into the closed instrument loop than did rapid operation. Automated value retrieval is recommended; we achieved a 3 to 5-fold time reduction for each injection cycle (ca. <2 min), and operator reading, recording, and digitization errors are eliminated.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。