7-Tesla MRI of the brain in a research subject with bilateral, total knee replacement implants: Case report and proposed safety guidelines

对一名接受双侧全膝关节置换术的研究对象进行7特斯拉脑部磁共振成像:病例报告及安全指南建议

阅读:1

Abstract

Recently, the first 7-T MR system was approved for clinical use in the United States. Unfortunately, relatively few metallic implants have undergone testing to determine if they are acceptable or pose hazards to research subjects and patients at this ultra-high-field strength. Therefore, in lieu of not performing a research or clinical MRI exam at 7-T, the supervising physician may make a decision to scan the individual with an untested metallic implant based on an analysis of the risks vs. the benefits. We present a case report of a research subject with bilateral, total knee replacement implants that safely underwent MRI of the brain at 7-T and provide guidelines for healthcare professionals to follow in order to ensure safety in research subjects or patients with metallic implants referred for 7-T scans.