Performance evaluation of a chemiluminescence microparticle immunoassay for CK-MB

CK-MB化学发光微粒免疫测定法的性能评价

阅读:1

Abstract

BACKGROUND: To verify and evaluate the performance characteristics of a creatine kinase phosphokinase isoenzymes MB (CK-MB) assay kit, which produced by Xiamen Innodx Biotech Co. Ltd. METHODS: Evaluation was carried out according to "Guidelines for principle of analysis performance evaluation of in vitro diagnostic reagent." The performance parameters included detection limit, linearity range, reportable range, recovery test, precision verification, interference test, cross-reactivity, matrix effect, and method comparison. RESULTS: The detection limit was 0.1 ng/mL. The assay had clinical linearity over range of 0.1 ng/mL-500 ng/mL. Reportable range was from 0.1 ng/mL to 1000 ng/mL. The average percent of recovery was 99.66%. The coefficient of variation (CV) for within-run and between-run of low CK-MB sample was 5.55% and 6.16%, respectively. As for high-level sample, it was 7.88% and 7.80%. In medical decision level, the relative deviation (Bias) of all interference tests was lower than 15%. When the sample had mild-hemolysis; hemoglobin ≤15 g/L; triglyceride ≤17 mmol/L; bilirubin ≤427.5 μmol/L; rheumatoid factor ≤206U/mL, there was no significant interference to be found. Moreover, assay kit had no cross-reaction with CK-MM and CK-BB. At last, total diagnostic accuracy of kit was 93.24%, when compared with refer kit. CONCLUSION: Overall the results of the verification study indicated the performance of kit is met the requirements of the clinical test.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。