Influence of technique used to attach the infusion set to peristaltic finger smart-pumps on dispensing time: an experimental study

将输液装置连接到蠕动指式智能泵的技术对输液时间的影响:一项实验研究

阅读:1

Abstract

BACKGROUND: Infusion sets designed for peristaltic finger smart pumps (PFSPs) are necessary for the pumps' accurate handling. We previously found that medication dispensing is occasionally incomplete following the calculated infusion time when using certain combinations of PFSPs and infusion sets at a Japanese hospital. Thus, in this study, we investigated the cause of this observed delay by determining the effect of infusion set attachment technique on dispensing time using a combination of three kinds of PFSPs and five kinds of polyvinyl chloride (PVC) and polybutadiene (PB) infusion sets. METHODS: PFSPs with their exclusive infusion sets were used. The PVC and PB infusion sets were either not stretched or stretched to 1-3 cm and attached to the PFSP's liquid delivery system. PFSP dispensing rates were set at 25-400 mL/h. The primary outcome was the time required to dispense 100 mL of saline in a volumetric flask. RESULTS: The complete dispensing time correlated with the input time for all equipment combinations when the infusion sets were not stretched before attachment to the PFSP (R(2) = 0.9998-1.0000). When stretched, the complete dispensing time was longer than the input time (P < 0.01-0.05, analysis of variance with Tukey-Kramer multiple comparisons). The maximum dispensing time extension ratio for the PVC and PB infusion sets was 141.8% and 113.0%, respectively. CONCLUSION: Certain attachment techniques for infusion sets can adversely prolong drug dispensing time. As such, pharmacists should provide medical staff with information about the devices used to administer drugs, as well as about the drugs themselves.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。