Analytical Performance of a Novel Latex Turbidimetric Immunoassay, "Nanopia TARC", for TARC/CCL17 Measurement: A Retrospective Observational Study

新型乳胶浊度免疫测定法“Nanopia TARC”检测TARC/CCL17的分析性能:一项回顾性观察研究

阅读:1

Abstract

Thymus- and activation-regulated chemokine (TARC, also known as CCL17) is used as a biomarker for atopic dermatitis. The methods currently used for its measurement are complex, time-consuming, and require large machinery, warranting the need for a method that is simple, has a quick turnaround time, and requires less complex machinery. We evaluated the analytical performance of a novel latex turbidimetric immunoassay method, "Nanopia TARC", on 174 residual serum samples from patients with skin or allergic diseases. This evaluation included the assessment of the limit of blank/detection/quantification (LOB/D/Q), precision, accuracy, linearity, interference, and commutability between Nanopia TARC and "HISCL TARC", based on the chemiluminescent enzyme immunoassay (CLEIA) method. The LOB/D/Q values were 13, 57, and 141 pg/mL, respectively. The coefficient of variation of the repeatability was 0.9-3.8%, and that of the intermediate precision was 2.1-5.4%. The total error of the accuracy was 1.9-13.4%. The linearity was 141 and 19,804 pg/mL for TARC. The correlation coefficient between Nanopia TARC and HISCL TARC determined using the Passing-Bablok regression analysis was 0.999. Furthermore, the concordance of diagnostic criteria with AD was 92%. Nanopia TARC was confirmed to have the same analytical performance for TARC measurement as the existing CLEIA method.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。