A Novel Screen-Printed Textile Interface for High-Density Electromyography Recording

一种用于高密度肌电图记录的新型丝网印刷纺织界面

阅读:1

Abstract

Recording electrical muscle activity using a dense matrix of detection points (high-density electromyography, EMG) is of interest in a range of different applications, from human-machine interfacing to rehabilitation and clinical assessment. The wider application of high-density EMG is, however, limited as the clinical interfaces are not convenient for practical use (e.g., require conductive gel/cream). In the present study, we describe a novel dry electrode (TEX) in which the matrix of sensing pads is screen printed on textile and then coated with a soft polymer to ensure good skin-electrode contact. To benchmark the novel solution, an identical electrode was produced using state-of-the-art technology (polyethylene terephthalate with hydrogel, PET) and a process that ensured a high-quality sample. The two electrodes were then compared in terms of signal quality as well as functional application. The tests showed that the signals collected using PET and TEX were characterised by similar spectra, magnitude, spatial distribution and signal-to-noise ratio. The electrodes were used by seven healthy subjects and an amputee participant to recognise seven hand gestures, leading to similar performance during offline analysis and online control. The comprehensive assessment, therefore, demonstrated that the proposed textile interface is an attractive solution for practical applications.