Comparative analysis of the quantitative point-of-care CareSTART™ total bilirubin with central laboratory total bilirubin assays in neonatal blood samples

对新生儿血样中即时定量CareSTART™总胆红素检测与中心实验室总胆红素检测进行比较分析

阅读:1

Abstract

This study aimed to assess and compare the accuracy of point-of-care CareSTART™ S1 Total Bilirubin test with a central laboratory total bilirubin assay using neonatal samples. This study was conducted using 152 paired measurements obtained from 122 neonates admitted to the neonatal intensive care unit. Total serum bilirubin (TSB) levels assayed with the central laboratory assay, laboratory bilirubinomter, trancutaneous bliribubin (TcB) instrument and CareSTART were compared using Bland-Altman analysis. The mean difference between the CareSTART and TSB values was -1.43 mg/dL and the 95% limit of agreement (LoA) was -4.25 to 1.39 mg/dL. CareSTART tended to underestimate total bilirubin concentrations compared with TcB, however, the LoA was narrower due to the smaller SD of mean difference for CareSTART. The CareSTART Total Bilirubin test provides an accurate alternative to TcB for total serum bilirubin measurement. Given its low-cost, ease-of-use, and portability, the use of CareSTART is expected to provide point-of-care measurements, especially in low-resource settings.