Performance Evaluation of Microscanner Plus, an Automated Image-Based Cell Counter, for Counting CD4+ T Lymphocytes in HIV Patients

Microscanner Plus(一种基于图像的自动细胞计数器)对 HIV 患者 CD4+ T 淋巴细胞计数的性能评估

阅读:5
作者:Woong Sik Jang, Junmin Lee, Seoyeon Park, Chae Seung Lim, Jeeyong Kim

Abstract

Counting CD4+ T lymphocytes using flow cytometry is a standard method for monitoring patients with HIV infections. Simpler and cheaper alternatives to flow cytometry are in high demand because getting access to flow cytometers is difficult or impossible in resource-limited settings. We evaluated the performance of the Microscanner Plus, a simple and automated image-based cell counter, in determining CD4 counts against a flow cytometer. CD4 count results of the Microscanner Plus and flow cytometer were compared using samples from 47 HIV-infected patients and 87 healthy individuals. All CV% for precision and reproducibility tests were less than 10%. The Microscanner Plus's lowest detectable CD4 count was determined to be 15.27 cells/µL of whole blood samples. The correlation coefficient (R) between Microscanner Plus and flow cytometry for CD4 counting in 134 clinical samples was very high, at 0.9906 (p < 0.0001). The automated Microscanner Plus showed acceptable analytical performance for counting CD4+ T lymphocytes and may be particularly useful for monitoring HIV patients in resource-limited settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。