Influence of dialysis membrane composition on plasma bisphenol A levels during online hemodiafiltration

透析膜成分对在线血液透析滤过过程中血浆双酚A水平的影响

阅读:5
作者:Sebastian Mas, Enrique Bosch-Panadero, Pedro Abaigar, Vanesa Camarero, Ignacio Mahillo, Esther Civantos, Didier Sanchez-Ospina, Alberto Ruiz-Priego, Jesus Egido, Alberto Ortiz, Emilio González-Parra

Conclusion

OL-HDF for 3 months with BPA-free dialyzer membranes was associated to a significant decrease in predialysis BPA levels when compared to baseline BPA levels while on a BPA-containing membrane.

Methods

A prospective study assessed plasma BPA levels in OL-HDF patients using BPA-free (polynephron) or BPA-containing (polysulfone) dialyzers in a crossover design with two arms, after a run-in OL-HDF period of at least 6 months with the same membrane: 31 patients on polynephron at baseline were switched to polysulfone membranes for 3 months (polynephron-to-polysulfone) and 29 patients on polysulfone were switched to polynephron for 3 months (polysulfone-to-polynephron).

Results

After a run-in OL-HDF period of at least 6 months with the same membrane, baseline pre-dialysis BPA was lower in patients on polynephron (8.79±7.97 ng/ml) than in those on polysulfone (23.42±20.38 ng/mL, p<0.01), but still higher than in healthy controls (<2 ng/mL). After 3 months of polynephron-to-polysulfone switch, BPA was unchanged (8.98±7.88 to 11.14±15.98 ng/mL, ns) while it decreased on the polysulfone-to-polynephron group (23.42±20.38 to 11.41±12.38 ng/mL, p<0.01).