Volumetric fluorescein angiography (vFA) by oblique scanning laser ophthalmoscopy in mouse retina at 200 B-scans per second

采用斜扫描激光检眼镜对小鼠视网膜进行容积荧光血管造影 (vFA),速度为每秒 200 次 B 扫描

阅读:9
作者:Weiye Song, Libo Zhou, Ji Yi

Abstract

Oblique scanning laser ophthalmoscopy (oSLO) is a recently developed technique to provide three-dimensional volumetric fluorescence imaging in retinas over a large field of view, without the need for depth sectioning. In this study, we present volumetric fluorescein angiography (vFA) at 200 B-scans per second in mouse retina in vivo by oSLO. By using a low-cost industrial CMOS camera, imaging speed was improved to 2 volumes per second, ∼10 times more than our previous results. Enabled by the volumetric imaging, we visualized hemodynamics at single capillary level in a depth-dependent manner, and provided methods to quantify capillary hematocrit, absolute capillary blood flow speed, and detection of capillary flow stagnancy and stalling at different vascular layers. The quantitative metrics for capillary hemodynamics enhanced by volumetric imaging can offer valuable insight into vision science and retinal pathologies.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。