High throughput evaluation of gamma-H2AX

γ-H2AX 的高通量评估

阅读:5
作者:Dane Avondoglio, Tamalee Scott, Whoon Jong Kil, Mary Sproull, Philip J Tofilon, Kevin Camphausen

Abstract

The DNA double-strand break (DSB) is the primary lethal lesion after therapeutic radiation. Thus, the development of assays to detect and to quantitate these lesions could have broad preclinical and clinical impact. Phosphorylation of histone H2AX to form gamma-H2AX is a known marker for irradiation-induced DNA DSBs. However, the first generation assay involves the use of immunofluorescent staining of gamma-H2AX foci. This assay is time consuming, operator dependent and is not scalable for high throughput assay development. Thus, we sought to develop a new assay using a high throughput electrochemiluminescent platform from Mesoscale Discovery Systems to quantify gamma-H2AX levels. The results show that our assay utilizes significantly less time and labor, has greater intra-assay reproducibility and has a greater dynamic range of gamma-H2AX versus irradiation dose.