Nano-biosupercapacitors enable autarkic sensor operation in blood

纳米生物超级电容器使血液中的传感器能够独立运行

阅读:18
作者:Yeji Lee, Vineeth Kumar Bandari, Zhe Li, Mariana Medina-Sánchez, Manfred F Maitz, Daniil Karnaushenko, Mikhail V Tsurkan, Dmitriy D Karnaushenko, Oliver G Schmidt

Abstract

Today's smallest energy storage devices for in-vivo applications are larger than 3 mm3 and lack the ability to continuously drive the complex functions of smart dust electronic and microrobotic systems. Here, we create a tubular biosupercapacitor occupying a mere volume of 1/1000 mm3 (=1 nanoliter), yet delivering up to 1.6 V in blood. The tubular geometry of this nano-biosupercapacitor provides efficient self-protection against external forces from pulsating blood or muscle contraction. Redox enzymes and living cells, naturally present in blood boost the performance of the device by 40% and help to solve the self-discharging problem persistently encountered by miniaturized supercapacitors. At full capacity, the nano-biosupercapacitors drive a complex integrated sensor system to measure the pH-value in blood. This demonstration opens up opportunities for next generation intravascular implants and microrobotic systems operating in hard-to-reach small spaces deep inside the human body.