A Surface Plasmon Resonance-based assay to measure serum concentrations of therapeutic antibodies and anti-drug antibodies

基于表面等离子体共振的检测方法测量治疗性抗体和抗药抗体的血清浓度

阅读:4
作者:Marten Beeg, Alessandro Nobili, Barbara Orsini, Francesca Rogai, Daniela Gilardi, Gionata Fiorino, Silvio Danese, Mario Salmona, Silvio Garattini, Marco Gobbi

Abstract

Therapeutic drug and immunogenicity monitoring (TDIM) is increasingly proposed to guide therapy with biologics, characterised by high inter-individual variability of their blood levels, to permit objective decisions for the management of non-responders and reduce unnecessary interventions with these expensive treatments. However, TDIM has not yet entered clinical practice partly because of uncertainties regarding the accuracy and precision of enzyme-linked immunosorbent assays (ELISA). Here we report the characterisation of a novel surface plasmon resonance (SPR)-based TDIM, applied to the measurement of serum concentrations of infliximab, an antibody against tumour necrosis factor α (anti-TNFα), and anti-infliximab antibodies. SPR has the obvious advantages of directly detecting and measuring serum antibodies in minutes, avoiding the long incubation/separation/washing/detection steps of the methods proposed so far, reducing complexity and variability. Moreover, drug and anti-drug antibodies can be measured simultaneously. This new method was validated for sensitivity and reproducibility, and showed cost-effectiveness over commercial ELISA kits. This method may be applied to other biotherapeutics. These data pave the way for the development of SPR-based point-of-care devices for rapid on-site analysis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。