Pragmatic Approaches to Reducing Radiation Dose in Brain Computed Tomography Scan using Scan Parameter Modification

使用扫描参数修改减少脑部计算机断层扫描辐射剂量的实用方法

阅读:7
作者:Mohammad Reza Choopani, Iraj Abedi, Fatemeh Dalvand

Background

High radiation dose of patients has become a concern in the computed tomography (CT) examinations. The

Conclusion

In this study, it was shown that in the brain CT scan imaging, the radiation dose was reduced by 30%-50% by increasing the slice thickness or changing the reconstruction filter. It is necessary to adjust the CT scan protocols according to clinical requirements or the special conditions of some patients while maintaining acceptable image quality.

Methods

The body mass measurement device phantom was repeatedly scanned by changing the scan parameters. To analyze the image quality, software-based and observer-based evaluations were employed. To study the effect of scan parameters such as slice thickness and reconstruction filter on image quality and radiation dose, the structural equation modeling was used.

Results

By changing the reconstruction filter from standard to soft and slice thickness from 2.5 mm to 5 mm, low-contrast resolution did not change significantly. In addition, by increasing the slice thickness and changing the reconstruction filter, the spatial resolution at different radiation conditions did not significantly differ from the standard irradiation conditions (P > 0.05).