Image Quality and Pulmonary Nodule Detectability at Low-dose Computed Tomography (low kVp and mAs): A phantom study

低剂量计算机断层扫描(低 kVp 和 mAs)的图像质量和肺结节可检测性:幻影研究

阅读:2
作者:Sepideh Iranmakani, Amir Reza Jahanshahi, Parinaz Mehnati, Tohid Mortezazadeh, Davood Khezerloo

Background

Nowadays, there has been a growing demand for low-dose computed tomography (LDCT) protocols. CT has a critical role in the management of the diagnosis chain of pulmonary disease, especially in lung cancer screening. There have been introduced several dose reduction

Conclusions

Here, we demonstrated that the LDCT scan has an insignificant effect on the perception of lung nodules. In this study, patient dose in lung CT was reduced by modifying of kVp and mAs about approximately 70%. Hence, to step in toward low-dose strategies in medical imaging clinics, using easy-implemented and feasible low-dose strategies may be helpful.

Methods

This study was designed in a two steps, in the first step, a feasible low-dose lung CT protocol was selected with quality assessment of accreditation phantom image. In the second step, the selected low-dose protocol with an appropriate image quality was performed on a manmade lung phantom in which there were objects equivalent to the pulmonary nodule. Finally, image quality parameters of the phantom at the appropriate scan protocol were compared with the standard protocol.

Results

A reduction of about 17% of kVp and 46% in tube current leads to dose reduction by about 70%. The contrast-to-noise ratio in the low-dose protocol remained almost unchanged. The signal-to-noise ratio in the low-dose protocol decreased by approximately 32%, and the noise level has increased by about 1.5 times. However, this reduction method hardly affected the detectability of nodules in man-made pulmonary phantom. Conclusions: Here, we demonstrated that the LDCT scan has an insignificant effect on the perception of lung nodules. In this study, patient dose in lung CT was reduced by modifying of kVp and mAs about approximately 70%. Hence, to step in toward low-dose strategies in medical imaging clinics, using easy-implemented and feasible low-dose strategies may be helpful.